5.2. aScope 4 Cysto specifications
Insertion portion
Bending angle
1
Insertion cord diameter
Distal end diameter
Maximum diameter
of insertion portion
Working length
Working channel
Minimum working
channel width
2
Storage and transportation
Transportation
temperature
Storage temperature
Relative humidity
Atmospheric pressure
Electrical power
Power requirement
LED power requirement
1.
Please be aware that the bending angle can be affected if the insertion cord is not kept
straight or have inserted endoscopic instruments.
2.
There is no guarantee that endoscopic instruments selected solely using this minimum
working channel width will be compatible in combination.
3.
Storage under higher temperatures may impact shelf life.
10
aScope 4 Cysto
210° ± 15°
,
min. 120 °
16.2 Fr ± 0.3 Fr /
5.4 mm ± 0.1 mm
(0.21" ± 0.004")
16.2 Fr ± 0.3 Fr /
5.4 mm ± 0.1 mm
(0.21" ± 0.004")
Max 18 Fr /
6.0 mm (0.24")
390 mm ± 10 mm
(15.4" ± 0.4")
Min. 6.6 Fr /
2.2 mm (0.086")
10 ~ 40 °C
(50 ~ 104 °F)
10 ~ 25 °C
3
(50 ~ 77 °F)
30 ~ 85 %
80 ~ 109 kPa
5 VDC 0.1 A input (from Ambu displaying unit)
18 mA (6.5 VDC) input (from Ambu displaying unit)
Optical system
Direction of view
Field of view
Depth of field
Illumination method
Sterilisation
Method of sterilisation
Operating environment
Temperature
Relative humidity
Atmospheric pressure
Altitude
aScope 4 Cysto
120° ± 10°
3 - 100 mm
LED
ETO
10 ~ 40 °C
(50 ~ 104 °F)
30 ~ 85 %
80 ~ 109 kPa
≤ 2000 m