Wall mount
Refer to illustration G.
Use appropriate fittings to attach the wall mount. For example, use the round holes or the slots of
the wall mount for screws or cable ties (not provided).
The RCM is only suitable for mounting at heights ≤ 2 m.
WARNING
Ensure that the wall mount is securely fixed in place.
1. Hold RCM at an angle and push it onto the wall mount. Adjust the angle to fit RCM in place.
2. Turn RCM clockwise until it clicks into place.
3. To remove RCM from the wall mount, turn it counterclockwise.
Sending data to AirView
RCM automatically sends the previous 24 hours of data to AirView once at approximately 12 pm
(based on the clock of the connected ventilation device), and once an hour after therapy has
stopped.
If automatic data transmission is missed or interrupted, it will resume when the connection is
re-established. No data will be lost.
To ensure daily automatic data transmission, ResMed recommends you to connect RCM once a day
for one hour, with the Vent. Input and Signal indicators on.
Note: Data more than seven calendar days old will not be sent to AirView.
Cleaning and maintenance
The exterior of RCM and the PSU can be cleaned with a damp cloth and an approved mild cleaning
solution.
The following cleaners and disinfectants are compatible for use when cleaning external surfaces of
RCM:
•
isopropyl alcohol*
•
bleach (1:10)* (may also be known as 'diluted hypochlorite').
*Suitable for weekly use
Always follow the manufacturer's recommended cleaning instructions.
WARNING
Ensure that the RCM and PSU are dry before reconnecting to the power outlet and ventilation
device.
Troubleshooting
Problem/possible cause
The Power indicator does not illuminate.
The Vent. Input indicator does not illuminate.
4
Solution
Check that the PSU is connected correctly to the power outlet and the
rear of RCM.
Check that the plug blade attachment is inserted into the PSU correctly.
Check that the PSU is the one provided with RCM.
Check that the ventilation device is turned on.
Check that the USB cable is connected correctly to the rear of RCM and