Introduction
Refer to illustration A.
ResMed Connectivity Module (RCM) provides cellular connection between a compatible ResMed
ventilation device and the ResMed AirView
RCM sends therapy and device data recorded in the ventilation device to the cloud-based AirView
system from home, wirelessly and automatically, to assist the remote display of patient data.
RCM also sends data to AirView on demand when requested via AirView (eg, for remote display and
troubleshooting).
Compatible devices
RCM is compatible with the following ventilation devices:
Astral
100/150
Stellar
100/130/150.
Intended use
RCM is intended to be used in the home environment, for the collection and transmission of
respiratory data to AirView. RCM will not control any clinical devices, nor provide interpretation of
data.
RCM is not intended for use on an aircraft.
General warnings and cautions
The following are general warnings and cautions. Specific warnings, cautions and notes appear with
the relevant instructions in the guide.
WARNING
• Only use the power supply unit and plug blade attachments provided with RCM.
• Beware of electrocution. Do not immerse RCM or any of its components in water. Always
unplug RCM before cleaning and make sure that all parts are dry before plugging it back in.
• Do not open or modify the device. There are no user serviceable parts inside. Repairs and
servicing should only be performed by an authorized ResMed service agent.
• The device should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the device should be observed to verify normal operation in the
configuration in which it will be used.
• The use of cables other than those specified for the device is not recommended. They may
result in increased emissions or decreased immunity of the device.
• Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm to any part of the device, including
cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
• The RCM device has not been tested or certified for use in the vicinity of X-ray, CT or MRI
equipment. Do not bring the RCM device within 4 m of X-ray or CT equipment. Never bring
the RCM device into an MR environment.
CAUTION
Do not use RCM outdoors.
Note: For any serious incidents that occur in relation to this device, these should be reported to
ResMed and the competent authority in your country.
system.
English
1