Aesculap
TPLO saw GT666
Issue
Saw blade cannot be
removed
10. Technical Service
Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
WARNING
►
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
►
Do not modify the product.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
►
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
16
Acculan 4
®
Cause
Distorted tool holder
Distorted saw blade saddle
shaft
Distorted clamp screw
Detection
Deformation
11. Accessories/Spare parts
Art. no.
GA675
GA346
GA678
GA679
GB020R
GB600
GB498R
GB487R
GB488R
GB489R
e.g.
GC538R
GC539R
GC544R - GC548R
TA004103
TA014554
Remedy
Have repaired by manufacturer
Replace saw blade
Replace clamp screw
Designation
Lid
Li-Ion battery short
Sterile funnel
Battery removal device
Hexagon wrench
STERILIT Power Systems oil spray
ECCOS fixation for straight machine
ECCOS fixation for lid
ECCOS fixation for sterile funnel
ECCOS fixation for battery removal device
TPLO saw blades
Fixing screw
Instructions for use for TPLO saw GT666 (A4 for
ring binder)