12. Technical data
12.1 Classification acc. to Directive 93/42/EEC
Art. no.
Designation
GT666
TPLO saw
12.2 Performance data, information about standards
Max. power
Max. oscillation frequency
Weight (operational)
Dimensions (L x W x H, ready for
operation)
Applied part
EMC
Conforming to standard
The product was subject to 350 preparation cycles as a test by the manu-
facturer and has passed.
12.3 Operating mode
Operating mode
12.4 Environmental conditions
Operation
Temperature
10 °C to 27 °C
Relative
30 % to 75 %
humidity
Atmospheric
700 hPa to 1 060 hPa
pressure
250 W
-1
15 000 min
1.44 kg ± 10 %
280 mm x 55 mm x 74 mm ± 5 %
Type BF
IEC/DIN EN 60601-1-2
IEC/DIN EN 60601-1
Operation with non-periodic load and
speed changes (type S9 pursuant to
IEC EN 60034-1)
■
30 second application, 30 second
pause
■
10 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Storage and transport
-10 °C to 50 °C
10 % to 90 %
500 hPa to 1 060 hPa
13. Disposal
Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
Class
Adhere to national regulations when disposing of or recy-
cling the product, its components and its packaging!
IIa
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufac-
turer as a free-of-charge service.
►
Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.
14. Distributor in the US/Contact in Can-
ada for product information and com-
plaints
Aesculap Inc.
3773 Corporate Parkway
Center Valley, PA, 18034,
USA
17