Turn the product 1 with the battery shaft 4 upwards.
►
Press both safety catches 9 on the lid 8 at the same time and remove
the lid 8.
►
Tightly grasp the battery shaft 4 at the lower end.
►
Tap the battery shaft 4 on the palm of the hand until the battery 7
slides out of the battery shaft 4 and can be removed.
6.1.5
Protection against inadvertent activation
To prevent inadvertent activation of the product during tool change, the
trigger can be locked.
To lock trigger:
►
Twist the safety catch 3 to position OFF.
The trigger 2 is blocked and the product 1 cannot be operated.
To unlock trigger:
►
Twist the safety catch 3 to position ON.
The trigger 2 is unlocked and the product 1 can be operated.
6.1.6
Attaching and detaching the saw blade
Risk of injury when attaching/removing sawblades
in the ON position through inadvertent activation
of the product!
WARNING
►
Only attach/remove tools in the OFF position.
Coupling the saw blade
►
Secure product 1 against inadvertent activation with the safety
catch 3.
►
Insert the TPLO saw blade 11 all the way into the tool holder 14 with
the connection side in the desired position,
►
Secure the TPLO saw blade 11 with the clamp screw 12 using the pin
spanner 13.
Uncoupling the saw blade
►
Secure product 1 against inadvertent activation with the safety
catch 3.
►
Loosen the clamp screw 12 with the pin spanner 13.
►
Pull out TPLO saw blade 1 from the tool holder 14.
6.2
Function checks
The function checks must be carried out prior to each use and after each
intraoperative battery change.
►
Make certain there is a battery 7 inserted.
►
Ensure that the lid 9 has completely clicked into place.
►
Pull on the TPLO saw blade 11 to check that the TPLO saw blade 11 is
securely coupled.
►
Check that the cutting edges of the TPLO saw blade 11 do not exhibit
any mechanical damage.
►
Release product 1 for use (ON position).
►
Briefly operate product 1 at maximum oscillation frequency.
►
Do not use the product 1 if it is damaged or defective.
►
Set aside the product 1 if it is damaged.
6.3
Safe operation
Coagulation of patient tissue or risk of burns for
patients and user through hot product!
►
Cool the tool during operation.
WARNING
►
Keep product/tool out of the reach of patients.
►
Let the product/tool cool down.
►
Use a cloth to protect against burns when
changing the tool.
Risk of infection or injury through aerosol forma-
tion or particles ejected from the tool!
►
Use suitable protection (such as waterproof pro-
WARNING
tective clothing, face mask, safety gases, suc-
tion).
Risk of injury and/or malfunction!
►
Always carry out a function check prior to using
WARNING
the product.
Risk of injury when using the product beyond the
field of view!
►
WARNING
Apply the product only under visual control.
Risk of injury and damage to the tool/system!
The tool may get caught in drapes (such as tex-
WARNING
tiles).
►
Do not let the tool come into contact with
drapes (such as textiles) during operation.
7