Automatic Cleaning/Disinfection With Manual Pre-Cleaning; Manual Pre-Cleaning With A Brush - Braun Aesculap Acculan 4 Mode D'emploi/Description Technique

Masquer les pouces Voir aussi pour Aesculap Acculan 4:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
7.7

Automatic cleaning/disinfection with manual pre-cleaning

Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
Note
The cleaning and disinfection machine used for processing must be ser-
viced and checked at regular intervals.
7.7.1

Manual pre-cleaning with a brush

Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
Mobilize non-rigid components during cleaning.
Thoroughly clean the product under running water.
Phase II
Mobilize non-rigid components during cleaning.
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold) -
-
RT (cold) -
-
Water
Chemistry/Note
quality
D–W
Until visually clean
D–W
Until visually clean
11

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Gt666

Table des Matières