pH or redox sensor holder /
gasket (in contact with the fluid)
pH or Redox probe of the 8202 refer to the related manual
Conductivity sensor of the
8222 (in contact with the fluid)
Pt1000 (in contact with the
fluid)
Electrodes of the 8222
• sensor C=1
• sensor C=0.1 or C=0.01
Fig. 7:
Materials used in the transmitter 8202 (without the probe)
or the 8222
6.3.2.
General technical data
Pipe diameter
Type of fitting
Nut between the 8202 and
the fitting
12
PVDF, stainless steel 1.4571
(316Ti) / EPDM
PVDF
stainless steel 1.4571 (316Ti)
• graphite
• stainless steel 1.4571 (316Ti)
DN25 to DN110 (DN15 to DN20
under specific conditions)
Adapter S022
G 1 1/2'' internal thread
English
Type 8202/8222
Technicaldata
Max. fluid temperature
The fluid temperature may be
restricted by the probe used (see the
related instruction manual), by the
pressure of the fluid and the material
of the adapter S022
See Fig. 2, Fig. 3, Fig. 4 and Fig. 5.
Max. fluid pressure
PN16
The fluid pressure may be restricted
by the probe used (see the related
instruction manual), by the tempe-
rature of the fluid and the material of
the adapter S022.
See Fig. 2, Fig. 3, Fig. 4 and Fig. 5.
ph measurement (8202)
• -2 to 16 pH or -580 to +580 mV
• Measurement range
• 0.001 pH or 0.1 mV
• Resolution
• ±0.02 pH or 0.5 mV
• Measuring error
• 0.5 pH unit or 30 mV (eg:
• Recommended min.
divergence of the pH
range associated to the
4-20 mA signal
range 6,7 to 7,2 pH or -20 mV
to +10 mV associated to the
4-20 mA output current)