Analogschnittstelle; Analog Interface; Interface Analogique - Ismatec MCP-CPF Mode D'emploi

Classe de protection ip 65
Table des Matières

Publicité

speed IN
direction
input 1
start
+26 V
remote
DC
input 2
GND
output 2
speed OUT
output 1
+5 V
speed intern
DC
Digitale Eingänge (TTL-Pegel)
Digital inputs (TTL-level)
Entrées numériques (niveau TTL)
Pin 2, remote
Pin 3, start
Pin 4, direction
Pin 6, input 1
Pin 8, input 2
Pin 13, speed intern
Analoge Eingänge
Analog inputs
Entrées analogiques
Pin 5, speed IN
0 – 5 V
/ 0 – 10 V
DC
DC
0 – 20 mA / 4 – 20 mA
Universal Ausgänge (PWM)
Universal outputs (PWM)
Sorties universelles (PWM)
Pin 14, output 1
Pin 15, output 2
Analog-Ausgang
Analog output
Sortie analogique
Pin 9, speed OUT
0-10 V
/ 0-7.2 kHz
DC
MCP-CPF PROCESS, 14-017, REV. C

Analogschnittstelle

Pin 1, GND (Masse)
Mit dem Schutzleiter verbundene
Masse. Bezugspotential für alle
anderen Eingänge.
Pin 2, remote
Für Umschaltung zwischen ma-
nueller Bedienung und der Analog-
schnittstelle. Zur Aktivierung der
Analog-Schnittstelle muss Pin 2 mit Pin
1 (GND) verbunden werden.
Pin 3, start
X
X
Im Normalbetrieb (Pin 2 offen)
dient Pin 3 als Fußschalter
(Grundeinstellungen Seite 17)
X
X
Im Remote-Betrieb (Pin 2 auf GND)
startet die Pumpe bei Verbindung mit
Pin 1 (GND)
Pin 4, direction
Wenn offen, dreht die Pumpe
im Uhrzeigersinn; wenn mit Pin
1 (GND) verbunden, dreht sie im
Gegenuhrzeigersinn.
Pin 5, speed IN
Externe Drehzahlsteuerung
(0 – 5 V, 0 – 10 V, 0 – 20 mA,
4 – 20 mA) Eingangsimpedanz und
Wahlmöglichkeiten mittels DIP switch
im Geräteinnern (siehe Seite 39)
Pin 6, input 1
Pin 8, input 2
Digitale Eingänge (TTL-Pegel). Aktiv,
wenn mit Pin 1 verbunden. Können
nur über RS232-Schnitt-stelle bedient
werden; z.B. mit einer Software
wie ProgEdit.
Pin 7, +26V
DC
Es stehen ca. +26 V
zur Verfügung
DC
(max. Strom 1A).

Analog interface

Pin 1, GND (ground)
Connected to the protective earth
conductor. Reference potential for all
other inputs.
Pin 2, remote
For changing between manual control
and analog interface. For activating
the analog interface, pin 2 must be
connected with pin 1 (GND).
Pin 3, start
X
X
In normal operation (pin 2 open)
pin 3 operates the foot-switch
(see basic settings Page 17)
X
X
In remote operation (pin 2 to GND)
the pump starts when connected to
pin 1 (GND)
Pin 4, direction
In the open position the pump turns
clockwise; when connected to pin 1
(GND) it turns counter-clockwise.
Pin 5, speed IN
External speed control
(0 – 5 V, 0 – 10 V, 0 – 20 mA, 4 – 20 mA)
Input impedance and input range can
be selected via a DIP switch inside the
pump (see Page 39)
Pin 6, input 1
Pin 8, input 2
Digital inputs (TTL-level). Active when
connected to pin 1. Are only retrievable
via RS232 interface; e.g. with a software
like the ProgEdit.
Pin 7, +26V
DC
About +26 V
are available
DC
(max. current 1A).

Interface analogique

Pin 1, GND (masse)
Connecté à la mise à terre. Potentiel de
référence pour toutes les autres entrées.
Pin 2, remote
Pour commuter du service manuel
à l'interface analogique.Pour activer
l'interface analogique, le pin 2 doit être
connecté au pin 1 (GND).
Pin 3, start
X
X
en exploitation normale (pin 2 ouvert),
le pin 3 sert d'interrupteur au pied
(réglages de base Page 17)
X
X
en exploitation à distance (pin 2 sur
GND), la pompe se met en route dès
qu'elle est connectée au pin 1 (GND)
Pin 4, direction
Si ouvert, le sens de rotation de la
pompe est celui des aiguilles d'une
montre; si relié avec le pin 1 (GND), la
pompe tourne dans le sens anti-horaire.
Pin 5, speed IN
Réglage externe du nombre de tours
(0 – 5 V, 0 – 10 V, 0 – 20 mA, 4 – 20 mA)
Impédance d'entrée et réglage de
zone au moyen de l'interrupteur DIP à
l'intérieur de l'appareil (v.p. 39)
Pin 6, input 1
Pin 8, input 2
Entrées numériques (niveau TTL). Actifs
si reliés au pin 1. Ne peuvent être
interrogés que par l'interface RS232, p.e.
avec le logiciel ProgEdit.
Pin 7, +26V
CC
Environ +26 V
sont à disposition
CC
(courant maximal 1A).
37 of 65