Tubing & Breast Pump Care; Tubing Care - Medela Symphony Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

7. Tubing & Breast Pump Care
Supplies needed:
• Mild dish soap
• Clean dish towel
• Clean sink or bowl
• Drinking quality water
• Damp (not wet cloth)
Parts needed:
• Tubing
• Symphony
®
Breast Pump
• Membrane cap
• Protective membrane
3
Remove, tubing, protective membrane,
and membrane cap. Clean protective
membrane and membrane cap by
following steps in Section 5.
Note
• Tubing should be washed if milk
is present or tubing is soiled.
• Cleaning tubing is not necessary
if condensation present is
from previous washings or
atmospheric conditions.
Warning
Breast pump kits are a personal
care item intended for use by a
single user only and should not be
shared between users.
18
7A. Tubing care
• Inspect tubing after each pumping session.
Caution
• Do not store wet or damp parts as mold may develop.
• Inspect the tubing for condensation or milk after each use.
Wash or replace the tubing immediately after use if soiled
or milk is observed in the tubing.
• If tubing becomes moldy, discontinue use and replace
tubing. Replacement parts can be purchased at
www.ShopMedela.com.
Note
If there is condensation in the tubing: Shake water
droplets out of the tubing and hang it up to dry or let the
pump run with tubing attached for 1–2 minutes or until dry.
4
Remove tubing from breast shield.
7
Use a damp (not wet) cloth to wipe the
pump and membrane area.

Publicité

Table des Matières
loading

Table des Matières