To Sterilize in the Hospital:
All disassembled parts can be autoclaved to a maximum of
272° F for 3 minutes at 29 PSIG or 250° F for 15-20 min-
utes at 15 PSIG . If autoclaved membrane caps may require
more force to attach to pump .
To avoid damage during and immediately after sterilization:
Sterilize parts without applying external pressure to
them .
Allow parts to cool before applying external pressure .
If gas sterilization is used, residual testing is recommended
to determine the proper length of aeration .
To Sanitize at Home:
Disassemble, refer to diagram on page 6 . Sanitize by boil-
ing for 10 minutes .

After each use:

Disassemble and wash all parts that come in contact
with the breast and milk in warm soapy water .
Rinse in clear water .
Air-dry on clean towel and cover parts when not in use .
All parts may also be washed in top rack of dishwasher .
For Quicker Cleaning:
Medela offers Quick Clean
kit disinfection .

Tubing Care:

Inspect tubing after each pumping session for condensa-
tion and or milk .