GB | INSTRUCTIONS FOR USE
Message displayed
Temperature increase inad-
missible
Temperature readings out of
range
Time out! Restart measure-
ment!
Ambient pressure out of
range
<Info> button: Permissible
ambient pressure range from
800 to 1100 mbar
OTHER MALFUNCTIONS |
POSSIBLE FAULTS
Malfunction Cause
Device can-
Battery is completely dis-
not be acti-
charged
vated
Device
Unfavourable operating con-
switches off
ditions (e.g. low humidity or
unsuitable floor coverings)
54
Cause
A spontaneous temperature
increase in the implant of
more than 2 K or above 39°C
Temperature data measured
from implant not plausible -
physiologically meaningless
data
The time between starting
measurement and success-
fully establishing commu-
nication was exceeded (60
seconds).
The permissible barometric
pressure during measure-
ment is too low or too high.
Error detection / troubleshooting
Connect the plug-in power unit. It takes
approx. 6 hours to charge the battery fully.
The Reader Unit can also be operated during
the charging process (with connected plug-in
power unit). Note: Charging is not possible at
an ambient temperature of more than 35°C.
Switch device back on. If the error persists,
contact Technical Service.
Reader Unit Set
Error detection /
troubleshooting
Measurement is stopped.
Leave to rest for 10 minutes.
If the error persists, contact
Technical Service.
Measurement is stopped.
Deactivate and reactivate
device. If the error persists,
contact Technical Service.
Restart measurement. Opti-
mize the distance between
antenna and the implant.
The Reader Unit can only be
used at barometric pressures
from 800 to 1100 hPa. Cur-
rent measurement is stopped.