GB | INSTRUCTIONS FOR USE
The <Up arrow> and <Down arrow> buttons
can be used to select the desired display con-
trast, which is confirmed with the <OK> but-
ton. Possible settings are 000-100 (in steps of
5).
CLEANING | DISINFECTION
WARNING
Risk of electric shock and fire!
Disconnect from mains before cleaning.
•
Do not use flammable or explosive deter-
•
gents or disinfectants.
Ensure that no fluids can penetrate into
•
the product; insert the SD card slot con-
nector into the Reader Unit.
NOTE
Mechanical cleaning or disinfection as well
as inappropriate detergents or disinfectants
can damage or destroy the product!
Only disinfect the product manually
•
Never sterilize the product
•
NOTE
Use detergents/disinfectants approved for
surface cleaning only as specified by
the manufacturer. Observe specifications
regarding concentration, temperature and
contact time.
48
Manual cleaning/disinfection
Wipe disinfection for electrical devices without
sterilization
Phase
I
Step
Wipe disinfection
T (°C/°F)
RT (room temperature)
≥1
t (min)
Conc. (%)
-
Water quality
-
Chemical
Meliseptol HBV wipes 50%
Propan-1-ol
Phase I
Remove any visible residue with a dis-
•
posable disinfectant wipe.
Wipe all surfaces of the visually clean
•
product with a fresh disposable disinfec-
tant wipe.
Observe the specified contact time (1
•
minute minimum).
Inspection
Check product for damages after each
•
cleaning/disinfection.
Immediately set aside product if dam-
•
aged.
Storage
Store products protected from dust in
•
a dry, dark and cool room with as little
bacterial contamination as possible.
Reader Unit Set