Cleaning | Disinfection - MIETHKE AAG FV907X Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 89
GB | INSTRUCTIONS FOR USE
The <Up arrow> and <Down arrow> buttons
can be used to select the desired display con-
trast, which is confirmed with the <OK> but-
ton. Possible settings are 000-100 (in steps of
5).

CLEANING | DISINFECTION

WARNING
Risk of electric shock and fire!
Disconnect from mains before cleaning.
Do not use flammable or explosive deter-
gents or disinfectants.
Ensure that no fluids can penetrate into
the product; insert the SD card slot con-
nector into the Reader Unit.
NOTE
Mechanical cleaning or disinfection as well
as inappropriate detergents or disinfectants
can damage or destroy the product!
Only disinfect the product manually
Never sterilize the product
NOTE
Use detergents/disinfectants approved for
surface cleaning only as specified by
the manufacturer. Observe specifications
regarding concentration, temperature and
contact time.
48
Manual cleaning/disinfection
Wipe disinfection for electrical devices without
sterilization
Phase
I
Step
Wipe disinfection
T (°C/°F)
RT (room temperature)
≥1
t (min)
Conc. (%)
-
Water quality
-
Chemical
Meliseptol HBV wipes 50%
Propan-1-ol
Phase I
Remove any visible residue with a dis-
posable disinfectant wipe.
Wipe all surfaces of the visually clean
product with a fresh disposable disinfec-
tant wipe.
Observe the specified contact time (1
minute minimum).
Inspection
Check product for damages after each
cleaning/disinfection.
Immediately set aside product if dam-
aged.
Storage
Store products protected from dust in
a dry, dark and cool room with as little
bacterial contamination as possible.
Reader Unit Set