7.3 Air Filter Maintenance
For an optimum operation of the device, the air filter can be cleaned by the user at least once
every two weeks or more often if this device is operated in a dusty environment. It is
recommended that the filter be replaced with a new one after 6 months of use; however,
depending on the air quality, the replacement time may vary.
CAUTION:
Dirty air filter may cause high operating temperatures that affect device
performance. Ensure the air filter is cleaned and fitted at all times.
1. Remove the dirty filter from the enclosure on the rear of the
device.
2. Wash the filter in warm water with a mild detergent, and rinse
with water. Allow the filter to air dry completely before
reinstalling. Do not use a filter that is not completely dry. If the
filter is torn, replace it.
3. Reinstall the filter.
NOTE: Please follow national requirements to dispose the unit properly.
8. Troubleshooting
The table below lists troubleshooting solutions for the problems that may happen. If the problem
persists, contact your equipment provider service agent.
P r o b l e m
P o s s i b l e C a u s e s
No display
1. The power cord is not
connected to the power socket.
2. LCD failure or controlled PCB
failure.
Display code incorrect
LCD failure or controlled PCB
failure.
Illuminant under LCD is not
LED failure
on
Buttons disable
Button failure
Air delivered is slow
1.
During ramp time.
2.
Filter is too dirty.
3.
Flow generator failure.
Data can not be copied to
1. SD card is full.
the SD card
2. SD card is not inserted
correctly.
3. Data on the SD card is
corrupted.
S o l u t i o n s
1. Ensure the power cord is
connected.
2. Contact your equipment provider
for repair.
Contact your equipment provider for
repair.
Contact your equipment provider for
repair.
Contact your equipment provider for
repair.
1. Check the ramp time setting
2. Change or clean the filter
regularly.
3. Contact your equipment provider
for repair.
1. Ensure the SD card has enough
capacity.
2. Ensure the SD card is inserted into
the slot.
3. Format the card.
14
Italiano
(11) Sveglia
Premere il pulsante "MENU" per selezionare il menù [Clock Alarm on/off]; premere il pulsante
"UP" o "DOWN" per impostare la sveglia su acceso o spento. Quando si imposta la sveglia, l'ora
compare sul lato sinistro del display. Premere il pulsante "UP" o "DOWN" per impostare l'ora
della sveglia. Una volta attivata la sveglia, premere il pulsante "START/STANDBY" per tacitare il
segnale acustico.
(12) Orologio
Premere il pulsante "MENU" per selezionare il menù [Clock XX:XX]; premere il pulsante "UP" o
"DOWN" per impostare l'ora attuale.
(13) Spegnere il dispositivo
Staccare la spina del cavo di alimentazione dalla presa elettrica e disinserire il cavo di
alimentazione dalla presa sul retro del dispositivo.
NOTA: Una volta confermata l'impostazione, premere il pulsante "MENU". Altrimenti, se non
viene effettuata alcuna operazione entro 20 secondi, il dispositivo torna
automaticamente in modalità standby senza salvare le modifiche.
(14) Indicazione eventi
Con il dispositivo in modalità standby, premere contemporaneamente i pulsanti "UP" e "DOWN"
per
visualizzare gli ultimi dati di Indice di Apnea (AI), Indice di Ipopnea (HI) e Indice di
Russamento (SI) sul display. Premere il pulsante "MENU" per visualizzare ogni indice in
sequenza. Per tornare alla modalità standby, premere il pulsante "START/STANDBY".
NOTA: Una volta riavviato il dispositivo, tutti gli indici vengono ricalcolati. Tutti i dati di questi
indici possono essere visualizzati unicamente dai terapisti della respirazione o dai
medici utilizzando il software di riproduzione Easy Compliance.
147