Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 109
iX Sense / English
items of the air tubing are subject to normal wear and tear and may eventually be replaced. Replace
the air tubing if they are damaged.
WARNING:
Do not use harsh soaps, detergents, solvents or alcohols to clean the tubing.
may reduce the life of the product.
WARNING: Do not wash or dry the air tubing at a temperature above 70°C (158°F)
WARNING: Do not use any cleaner containing fragrance or conditioners as they will leave a residue.

Air Filter

5.3
For an optimal operation of the device, the Ultra filter is disposable and should be replaced after 6 months
of use or sooner if it appears dirty; however, depending on the air quality, the replacement time may vary.
1.
Open back cover and remove the dirty filter from the enclosure on the rear of the device.
2.
Replace a new filter and close back cover.
WARNING: The Ultra filter is Not washable.
WARNING: Only APEX-supplied filters should be used as replacement filters.
CAUTION: Dirty air filter may cause high operating temperatures that affect
device performance.
18
These solutions

Publicité

Table des Matières
loading

Table des Matières