4.6.8
Installation dependent on the temperature of the
medium being measured
The mounting position of the sensor is dependent on the
temperature of the medium being measured T
aware of the following mounting options!
Fig. 13: Installation at T
medium
Fig. 14: Installation at T
medium
18 - EN CI/FCB300/FCH300-X1 Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350
. Be
medium
-50°... 120 °C (-58 ... 248 °F)
-50°... 200 °C (-58 ... 392 °F)
4.6.9
Installation with option TE1 "extended tower
length"
IMPORTANT (NOTE)
The sensor must only be insulated in conjunction with the TE1
"Extended tower length" option, as shown in Fig. 15.
Fig. 15: Installation at T
medium
1 Insulation
4.6.10 Notes about EHEDG conformity
WARNING – Risk of poisoning!
Bacteria and chemical substances can
contaminate or pollute pipeline systems and the
materials they are made of.
Observe the following instructions in installations
conforming to EHEDG requirements.
- EHEDG certification requires a self draining installation, only
possible with a vertical installation.
- In order to achieve compliance with EHEDG requirements,
the combination of process connection and gaskets
selected by the operator must consist solely of EHEDG-
compliant parts. Note the information in the latest version of
the following document: EHEDG Position Paper: "Hygienic
process connections to use with hygienic components and
equipment".
-50°... 200 ℃ ( -58 ... 392 °F)