1.
Wipe all surfaces of the chip and wireless receiver with one of the following:
a. 70% Isopropyl alcohol
b. 10% (1:10) chlorine bleach to water solution
c. Quaternary ammonium chloride solution
2.
Inspect for visible debris and repeat the above cleaning step as needed.
3.
Dry cleaned parts before use.
CAUTIONS:
•
To avoid permanent damage to the chip and wireless receiver, do not use undiluted
bleach (5% - 5.25% sodium hypochlorite) or any other cleaning solution not
recommended.
•
Do not immerse the chip and wireless receiver in any liquid solution.
•
Do not sterilize by irradiation, steam, autoclave or ethylene oxide.
Customer Support
For answers to frequently asked questions (FAQ's) and product support, along with
troubleshooting for your Masimo SafetyNet Alert product, please go to
www.mymasimo.co.uk/safetynet-alert or email customercare@masimo.com.