Analog Load Cell; Jumper Settings; Connections - Mettler Toledo IND780 Guide D'installation

Masquer les pouces Voir aussi pour IND780:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 295
2.7.2.

Analog Load Cell

2.7.2.1.

Jumper Settings

One jumper (W1) on the IND780 analog scale option board permits the gain of the analog section
to be set to either 2 mV/V or 3 mV/V. The factory default is 3 mV/V, which normally will work well
for both 2 mV/V and 3 mV/V load cells. If 2 mV/V load cells are used, the jumper can be changed
to the 2 mV/V position. Figure 2-40 shows the jumper's location and settings on the board.
The 2 mV/3 mV jumper settings are described in Table 2-6.
Jumper
2.7.2.2.

Connections

TO AVOID DAMAGE TO THE PCB OR LOAD CELL, REMOVE POWER FROM THE IND780 TERMINAL AND WAIT AT
LEAST 30 SECONDS BEFORE CONNECTING OR DISCONNECTING ANY HARNESS.
When using the IND780 with one or more analog load cells, load cell connections are made to the
connector located on the Analog Load Cell option board. The 7-position connector is visible at top
64057253
|
14
|
08/2014
Figure 2-39: Analog Load Cell Option Board
Figure 2-40: Analog Load Cell Board Millivolt Jumper Location and Settings
On
Off
X
When the jumper is closed, the connected scale's load cell setting is 2 mV/V.
W1
X
When the jumper is open, the connected scale's load cell setting is 3 mV/V
Gain Setting Jumper
Table 2-6: Millivolt Jumper Description
Description
NOTICE
METTLER TOLEDO IND780 Installation Guide
Pin 1
2-25

Publicité

Table des Matières
loading

Table des Matières