PDX Home Run Cable Grounding and Shielding

Ground and shield terminations are a critical part of the system's immunity to noise and electrical
surges. The following procedures describe proper cabling methods for the harsh and panel
enclosures, respectively.
2.7.4.3.1.
Cable Preparation, Both Enclosure Types
1. Use metal shears to cut back the braided outer metal sheath, about 8" from the end, then trim
2. Push the sheath about 5" back down the cable, so that it compresses as shown in Figure
3. Push the sheath down over the tape to create a neat edge.
2.7.4.3.2.
Cable Installation, Harsh Enclosure
1. Pass the cable through the body of the gland, and slide the gland down to meet the sheath. If
64057253
|
14
|
08/2014
the cut edge of the sheath to make a neat edge. Take care not to cut the external drain wire
when trimming the sheath.
2-52. Use electrical tape to keep the sheath in position.
Figure 2-52: Outer Sheath Secured with Electrical tape
Figure 2-53: Outer Sheath Pushed Up Over Taped End
necessary, loosen the clamp's screw to allow the sheath to fit inside it. Note that the sheath is
completely under the nut and against the body of the gland, and that the external drain wire
emerges between the two pieces of the clamp.
Outer drain wire
Figure 2-51: Outer Sheath Trimmed
METTLER TOLEDO IND780 Installation Guide
2-35