4
5
5.1
04/13
Order number 72196694E
Cleaning
Clean the weighing platform with a soft cloth soaked
➜
with a mild cleaning agent.
Take off the load plate and remove any dirt and foreign
➜
substances which may have collected underneath it.
Do not use any hard objects to do so. Do not open the
weighing platform.
Technical Data
Technical Data of the Weighing Cell
PBA220-...
except PBA220-CC600
Sensitivity
2 mV/V +/– 0.2 mV/V
Input resistance
410 Ω +/– 10 Ω
Output resistance 350 Ω +/– 4 Ω
Supply voltage
Recommended:
5–15 V DC/AC
Maximum: 20 V DC/AC
OIML approval
Cell type MT1022: TC 5442 Cell type IL1000: TC 6042
Cell type MT1241: TC 5382
Cell type MT1260: TC 5367
PBA220-CC600
2 mV/V +/– 0.002 mV/V
380 Ω +/– 15 Ω
350 Ω +/– 1 Ω
Recommended:
5–15 V DC/AC
Maximum: 20 V DC/AC
User Manual PBA220
METTLER TOLEDO
11