Connection Configurations; Connection B - Fore-Vacuum Pump - Agilent Technologies TV 301 Mode D'emploi

Table des Matières

Publicité

17
Technical Information
TV 301Controller Installation

Connection Configurations

Figure 55

Connection B - Fore-Vacuum Pump

A flange KF 16 NW is available to connect the Turbo-V301 pump to
the forevacuum pump. A hose or vacuum approved pipe can be used.
If a rigid pipe is used, any vibration generated by the mechanical
pump must be eliminated through the use of bellows.
The Turbo-V 301 pump is characterized by its high compression ratio also for oil
NOTE
vapors. When using a mechanical oil-sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore-vacuum pump in order to
prevent oil back-streaming.
272/284
TV 301 Navigator User Manual / 87-900-946-01 (N)

Publicité

Table des Matières
loading

Table des Matières