Chapter 5,
Calibration

Calibrating the Sirius Multigas Detector

Each Sirius Multigas Detector is equipped with an Autocalibration
feature to make unit calibration as easy as possible.
The Autocalibration sequence resets instrument zeroes and adjusts
sensor calibration for known concentrations of calibration gases.
Table 5-1. Autocalibration and Required Calibration Cylinders
SENSORS
EXPECTED
GAS*
CONCENTRATION
Combustible
58% LEL
Oxygen
15%
Carbon Monoxide
60 ppm
Hydrogen Sulfide
20 ppm
VOC
100 ppm isobutylene
*Factory Default
NOTE:
Refer to Chapter 4, Setting up the Multigas Detector, for instructions on changing the
Autocalibration expected gas concentrations if calibration gas with concentrations other
than those listed above will be used to calibrate the instrument.
The expected gas concentrations must match the gas con-
centrations listed on the calibration cylinder(s). Failure to
follow this warning will cause an incorrect calibration, which
can result in serious personal injury or death.
FOUR-GAS
CYLINDER
(P/N 10045035)

" WARNING

"
ISOBUTYLENE
(P/N 10028038))
5-1