The Sirius Multigas Detector is equipped with an optional green "SAFE"
LED which flashes every 15 seconds under the following conditions
The green SAFE LED is enabled
Instrument is on the normal Measure Gases page
Combustible reading is 0% LEL or 0%CH 4
Oxygen (O 2 ) reading is 20.8%
Carbon Monoxide (CO) reading is 0 ppm
Hydrogen Sulfide (H 2 S) reading is 0 ppm
VOC reading is 0 ppm
No gas alarms are present (low or high)
Instrument is not in Low Battery warning or alarm
CO, H 2 S, VOC, STEL and TWA readings are 0 ppm.

Operating Beep

The Sirius Multigas Detector is equipped with an optional operating
beep. This operating beep activates every 30 seconds by momentarily
beeping the horn and flashing the alarm LEDs under the following
conditions:
Operating beep is enabled
Instrument is on normal Measure Gases page
Instrument is not in Battery warning
Instrument is not in Gas alarm.
Figure 3-8. Heartbeat