Changing Response Factor
To change your current Response Factor, press RESET/
Response Factor Page.
•
Up and down arrows appear on the display.
•
User can now scroll, using the PAGE/
•
At any time, user can select the option display by pressing the
ON-OFF/ACCEPT button.
•
The first five Response Factors in the list are called favorites
(these can be set using our MSA FiveStar Link program).
•
The user has the option to turn OFF the PID (VOC detection), if
desired.
•
If the gas of interest isn't in the favorites list, select -MORE- to
scroll through the entire list of pre-programmed Response Factors
in alphabetical order.
Selecting a Custom Response Factor
If the gas of interest in is not in the pre-programmed list, the user can
use a Custom Response Factor if the multiplier is known for the gas
compared to the Isobutylene calibration gas. To do this:
1. Go to the Response Factor page and press RESET/
2. Scroll to and select -CUSTOM-.
3. Enter your desired eight-character identifier and multiplier.