Cal Lockout Enable:
•
To disable calibration, turn this feature ON
•
When ON, calibration is only accessible via the Setup
mode and password (if enabled)
•
CAL Due Alert
•
To disable CAL Due messages, turn this feature OFF.
•
When ON, the number of days (1 to 180) between
calibrations can be set. User will be required to
acknowledge an overdue calibration at turn ON.
•
Warm Up Info:
•
Turning this selection OFF causes the instrument NOT to
display alarm setpoints at power-ON
•
Time (if data logging option installed)
•
Date (if data logging option installed)
7. LEL/CH 4 Setup
•
Sensor ON/OFF (turns the sensor ON or OFF)
•
Display Combustible Gas Type?
•
Methane
•
Pentane
•
Hydrogen
•
Propane
•
LEL or Ch 4 mode (displays % LEL (for any gas)
or % CH 4 (for Methane only)
•
Low Alarm (sets the low combustible alarm)
•
High Alarm (sets the high combustible alarm)
•
Cal Gas (sets the expected combustible calibration gas)
8. O 2 Setup
•
Sensor ON/OFF (turns the sensor ON or OFF)
•
Low Alarm
•
High Alarm
9. CO Setup
•
Sensor ON/OFF (turns the sensor ON or OFF)
4-7