4. Push the ON-OFF/ACCEPT button to zero the instrument.
•
Instrument must be in fresh air to perform the zero.
•
CAL ZERO flashes.
NOTE: To skip the Zero procedure and move directly to the
calibration span procedure, push the RESET/
If no button is pushed for 30 seconds, the instrument
returns to the Measure mode.
•
Once the zeros are set, CAL SPAN? flashes (FIGURE 5-3).
Figure 5-2a. Zero Flag
Figure 5-2b. Zero Flag
button.