Mounting of
Multidoser (optional)
RotoCom Memory Unit
(optional)
Setting the Communication
18
RotoPol-31/-35
Instruction Manual
RotoPol-31/-35, combined with a specimen mover, can be equipped
with Multidoser for automatic dosing of suspensions and
lubricants. To connect Multidoser, see the appropriate Instruction
Manual.
RotoPol-31/-35, combined with RotoForce-3 or RotoForce-4 and
Multidoser, can be equipped with a RotoCom memory unit. To
connect RotoCom to RotoPol, see the appropriate Instruction
Manual.
m
Press MENU
COMMUNICATION
STANDARD
D
U
L
L
D
U
COMMUNICATION
ROTOCOM
o
Press STOP
m
MENU
repeatedly until the operation menu appears again.
to activate the set-up menu.
Move to the COMMUNICATION menu
using the MENU
Press SELECT L to toggle between
STANDARD and ROTOCOM:
- Select STANDARD if RotoPol is used as
a single unit or if combined with a
RotoForce-3/-4 and/or Multidoser.
- Select ROTOCOM when RotoPol is
connected to a RotoCom unit.
when the desired setting appears, or press
m
keys.