Medela Invia Motion Mode D'emploi page 23

Masquer les pouces Voir aussi pour Invia Motion:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 67
Battery fully discharged
Change Pressure
On/Off
Setting
Constant/
Mute/
Intermittent Mode
Info Used Time
Info used time
The Invia Motion NPWT pumps have a runtime of 15 or 60 days. After one hour of use, the timer is activated
and will continue even when the pump is turned off. During the first hour of use, the Invia Motion NPWT
pump can be turned on and off without starting the timer. If the timer function is used to readout the therapy
time, one hour needs to be added to the displayed runtime. Press [
operation time.
The example below displays 6 days and 23 hours of operation time used. The remaining runtime of the Invia
Motion – 15 Days version would therefore be 8 days and 1 hour.
On/Off
Change Pressure
Setting
Constant/
Mute/
Intermittent Mode
Info Used Time
Runtime expired
After 15 or 60 days, a blinking symbol on the display alike [
NPWT pump.
Change Pressure
Setting
Constant/
Intermittent Mode
Info Used Time
Example for 15 days version (blinking)
The Invia Motion NPWT pump runtime versions of 15 or 60 days provide an additional safety usage time
of 72 hours. During this time the device can be operated as usual. An acoustic signal will occur after
72 hours and the negative pressure will not be maintained.
Troubleshooting:
Charge battery
The battery symbol [
sounds and the negative pressure is not maintained. Connect the pump
to the power supply. When the pump is not connected to the power
supply within 10 minutes, it switches off.
On/Off
Mute/
] blinks on the display, an acoustic signal
] for information about the used
] will indicate the expiration of the Invia Motion
23

Publicité

Table des Matières
loading

Table des Matières