Battery low
Approximately 30 minutes before the battery is fully discharged, an acoustic signal will sound and the battery
symbol [
] will start to blink. The acoustic signal can be muted, but the battery symbol continues to blink
until the battery is fully discharged. The set pressure is maintained during the 30 minutes, but the battery
should be recharged as soon as possible. Fully charge battery. Charging time can be up to 4 hours.
Battery fully discharged
The battery symbol [
] blinks on the display, an acoustic signal sounds and the negative pressure is not
maintained. Connect the pump to the power supply. When the pump is not connected to the power supply
within 10 minutes, it switches off.
Connecting power source
Plug the Invia Motion power supply into a wall outlet and open the small protective cover to access the DC
socket. Plug the coaxial power supply into the DC socket.
Moving bars on the battery display will indicate charging is occurring [
].
1. Battery is fully
2. Battery is being charged
3. Battery is fully charged
discharged
(moving bars)
Disconnecting power source
When the battery is fully charged (after approximately 4 hours) the following non-blinking symbol will appear on
the display
. Charging of the battery can be slowed when exposed to high temperatures. To disconnect
the Invia Motion NPWT pump from the electrical source, hold the end of the plug and remove it from the pump.
Close the DC socket by pressing the small protective cover over the opening.
20