Operating instructions
Invia Motion NPWT pump settings
Change pressure level
Press [
] and hold for 3 seconds to change the pressure level. The set pressure value symbol on the
display will start to blink. You now have 5 seconds to press [
pump remains at the previous setting. Each time the button is pressed the next available pressure will
appear. When the appropriate pressure level appears on display, the pump will adopt it automatically after 5
seconds. The pressure may only be changed if instructed by a physician or qualified healthcare professional.
Available pressure levels
mmHg
Change therapy mode
The Invia Motion NPWT pump is able to deliver constant or intermittent ( 5 min. on / 2 min. off ) suction.
Constant mode
Intermittent mode
Press [
] and hold for 3 seconds to change the therapy mode. The therapy mode symbol on the display will
start to blink. You now have 5 seconds to press [
previous mode. Each time the button is pressed the next available therapy mode will appear. When the
appropriate mode symbol appears, the pump will adopt it automatically after 5 seconds. The therapy mode may
only be changed if instructed by a physician or a qualified healthcare professional.
Change Pressure
On/Off
Setting
Constant/
Mute/
Intermittent Mode
Info Used Time
Pump in constant mode
Mute / Information
Press [
] to obtain information about the runtime used or to mute the acoustic signal.
-40
-60
-80
-100
] to change the mode. Otherwise the pump remains at the
Change Pressure
On/Off
Setting
Constant/
Mute/
Intermittent Mode
Info Used Time
Pump in intermittent mode (on)
] for a new pressure setting, otherwise the
-125
-150
-175
Change Pressure
Setting
Constant/
Intermittent Mode
Info Used Time
Pump in intermittent mode (off)
On/Off
Mute/
17