4. Connect the dressing tubing to the pump tubing.
Make sure both parts are aligned correctly as shown in the picture below. Push together until you hear
a click.
For dressing applications, please refer to the instructions for use
provided with the Invia dressing kits.
Switch Invia Motion NPWT pump on
1. Press [
] to switch the Invia Motion NPWT pump on.
2. When turned on for the first time, the Invia Motion NPWT pump will start up in the default factory setting
(constant mode with pressure level of –125mmHg).
Change Pressure
Setting
Constant/
Intermittent Mode
Info Used Time
If the Invia Motion NPWT pump is turned off, the setting will be as previously set when the pump is turned
back on.
Run-time, timer and safety time
087.4022 Invia Motion – 15 Days Negative Pressure Wound Therapy system
087.4024 Invia Motion – 60 Days Negative Pressure Wound Therapy system
These two Invia Motion NPWT pump versions have a runtime of 15 and 60 days. After one hour of use, the timer
is activated and will continue even when the pump is turned off. During the first hour of use, the Invia Motion
NPWT pump can be turned on and off without starting the timer. If the timer function is used to readout the
therapy time, one hour needs to be added to the displayed runtime. Press [
operation time.
After 15 or 60 days, a blinking symbol on the display alike [
NPWT pump.
Change Pressure
Setting
Constant/
Intermittent Mode
Info Used Time
Example for 15 days version (blinking)
On/Off
Mute/
On/Off
Mute/
] for information about the used
] will indicate the expiration of the Invia Motion
15