Composition And Characteristics; Operating Instructions - Medela Invia White Foam NPWT Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15

Composition and characteristics

The Invia White Foam is made from 100% polyvinyl alcohol.
The Invia White Foam is a hydrophilic, enormously absorbent foam. This means it is ideal
for the effective removal of wound exudate and tissue fragments.
Combined with the Invia Negative Pressure Wound (NPWT) Therapy Systems with negative
pressure levels between 80 mmHg and 150 mmHg, Invia White Foam ensures reliable re-
moval of wound exudate from the wound. The applied negative pressure is distributed
evenly on the base of the wound.
An intact sealing system guarantees a reliable airtight closure of the wound and thus an
optimal use of negative pressure and exudate transport.

Operating instructions

– Invia White Foam must only be used by medical and nursing staff.
– Take Invia White Foam from undamaged packaging.
– If the packaging is damaged, you must not use Invia White Foam.
– Follow all the usual guidelines for sterile treatment when applying Invia White Foam.
– Always cut Invia White Foam with sterile instruments to the desired size.
– Make sure that no scraps come into the wound bed when cutting Invia White Foam.
– Injuries of the deep tissue can be filled with several layers of Invia White Foam.
– If necessary, Invia White Foam can be used with appropriate antiseptics, either prior to
fitting or using installation during the negative pressure therapy.
– Record the number of Invia White Foam pieces brought in the wound cavity in
the patient's chart.
– Do not apply alcoholic solutions, as this compromises the integrity of Invia White Foam
and its functioning.
– Make sure that Invia White Foam does not cover the skin around the wound or
the growing epithelium.
– If there is evidence of an infection of the wound, Invia White Foam must be changed reg-
ularly, at least once every 24 hours until the signs of the infection have gone down signifi-
cantly or disappeared altogether.
– An intermediate, non-stick wound cushion (limiting membrane) must be used when Invia
White Foam comes in close contact with (parts of) blood vessels, tendons, nerves and
other vulnerable vital organs and tissues.
– Always look for unexpected blood loss and instruct the patient accordingly.
– In the treatment of sternum wounds, controlling the blood pressure and ventilation
is mandatory.
– Invia White Foam must be changed regularly within a maximum of 72 hours.
– If you see any signs of dehydration and adherence, Invia White Foam must be rehydrated
with Ringer's solution first before it can be removed from the wound bed.
– Never use previously used Invia White Foam again.
– Discard Invia White Foam in accordance with local regulations.
5

Publicité

Table des Matières
loading

Table des Matières