Fig. 78: To enter the reference value of the test solution, tap the button (1)
1.
Start the calibration routine you want.
Device menu è Calibration è Select universal input è Zero point calibration
2.
Clean the sensor and dip it in the test solution. Wait for a stable reading display, and
confirm the measurement result with "OK" (See Fig. 77.)
3.
Enter the reference value of the test solution. Confirm with "OK" (See Fig. 78.)
4.
The system generates a summary log of the determined calibration values.
Acknowledge the log with "OK".
A failed calibration can be aborted at this point and discarded.
5.
Select "Yes" to assume the determined calibration values and log the calibration.
Select "No" to discard them.