TROUBLESHOOTING

Symptom
Water comes into the powder chamber
or leaks from the cap
No powder/air jet coming from the unit •
No water spray is coming from the
nozzle
Water leakage between the
handpiece and the nozzle
(unusual leak of droplets)
Water leakage between the nose
connection and handpiece
Air and/or powder is leaking from the
threads of the cap
84
FB-543_HS_rev_A_01_Handy 3 Perio & Handy 3 PLUS.indd 84
Solutions
Check the dental unit connection
Check the state of the o-rings of the turbine connection
Connect the unit to the dental unit (be careful, the powder chamber must be empty)
Dry and fill the powder chamber
Send the unit to an approved Hu-Friedy & EMS repair center
Cut immediately the air input by releasing the foot pedal
Wait for 1-2 minutes for the system to decompress
Disconnect the unit from the dental unit
Press the foot control pedal of the dental unit
if air comes out of the turbine connector, it is your dental unit that causes the problem
If air comes out, it is the device that causes the problem
Unscrew the cap above the wash-hand basin, because the powder can be ejected. A sealed apparatus, even
disconnected, can indeed remain under pressure
Empty the powder chamber and screw the cap back on
Connect the unit to the dental unit (be careful, the powder chamber must be empty)
Connect the handpiece to the unit
Press the foot control pedal of the dental unit
If air comes through the handpiece connector, the handpiece is blocked. Clean the handpiece
If no air comes out of the handpiece connector, the unit is blocked
Send the unit to an approved Hu-Friedy & EMS repair center
Change the PERIO-FLOW Nozzle
Remove the nozzle and verify that a water jet comes out from the handpiece
Send the device to a certified Hu-Friedy & EMS Repair Center
Check the position of the PERIO-FLOW Nozzle and its position on the handpiece
Change the PERIO-FLOW Nozzle
Check for residues between the PERIO-FLOW Nozzle and the handpiece
Check that nothing affects the alignment of the PERIO-FLOW Nozzle
Send the device to a certified Hu-Friedy & EMS Repair Center
Check the O-rings on the nose connection of the device, replace them if necessary
Change the handpiece
Send the device to a certified Hu-Friedy & EMS Repair Center
Check the seal and the cleanliness of the thread on the powder chamber and on the cap
Replace the seal or the cap if necessary
Send the device to a certified Hu-Friedy & EMS Repair Center