Checking the turbine connection (Typical is shown)

The unit is equipped with an adapter specially designed for the turbine connection of your dental
unit. Only use the unit with this special turbine connection. To connect it to another type of turbine will
damage it.
The turbine of your dental unit must not be under pressure when to connect the unit. Do not activate
the turbine footpedal. If you have a turbine equipped with light, turn the light off.
Make sure the o-rings of your turbine connection are in good condition. A turbine connection with
o-rings in bad condition can damage the unit.
The user should check the compatibility of this connection with his chair.