Cleaning the handpiece

Cleaning steps:
• Remove the PERIO-FLOW Nozzle from handpieces.
• Remove any externally attached soiling by brushing carefully with a soft surface brush or cloth.
• Soak all parts in the cleaning solution for, as a minimum, the time and the concentration specified by
the manufacturer of the cleaning agent (ENZOL, 2 %, 10 min).
• Brush all lumens and surfaces with a suitable lumen / surface brush until all visible soiling is removed.
• Rinse through all product lumina (e.g. irrigation and aspiration connection) with the cleaning solution
(ENZOL, 2 %), at least 5 times in the flow direction (no back rinsing), using a disposable syringe (min.
volume 50 ml)
• Submerge all components completely in deionized water.
• Rinse through all product lumina (e.g. irrigation and aspiration connection) with deionized water at
least 3 times in the flow direction (no back rinsing), using a disposable syringe (min. volume 50 ml)
and the Easy Clean.
• Repeat the cleaning process if the last rinse does not run clear, or if stains are still visible on the
product.
• Dry all components at room temperature.
• Place the handpiece in a sterilization bag.
58