1. Insert the PERIO-FLOW Nozzle into the periodontal pocket. The penetration depth depends on the
pocket depth. Do not force the nozzle into the pocket because this could hurt soft tissues.
2. When the PERIO-FLOW Nozzle is inserted in the pocket, press the pedal of your device for 5 seconds.
3. Place the PERIO-FLOW Nozzle inside the pocket for 5 seconds maximum per site. Repeated insertion
of the nozzle in the same site is not recommended. Carry on the treatment with short tiny vertical motions
of the PERIO-FLOW Nozzle.
Hu-Friedy & EMS powder/air mixture confined by a water jet is sprayed into the periodontal pocket. Due
to the mechanical action of the air, powder and water spray, the subgingival biofilm is eliminated from
the root surface.