CLEANING AND STERILIZING
Unclogging / Cleaning the handpiece using the Easy Clean
Clean the handpiece with the Easy Clean supplied with the unit before reprocessing.
Rinse through central lumina in the normal direction of flow (no back rinsing) using a 2 ml
disposable syringe filled with drink water applied to the purge adaptor. Ideally use distilled or deionized
water is used for all rinsing steps.
Blow compressed air inside the handpiece to dry it.
Clean the handpiece before sterilizing. (see the "Cleaning the handpiece" section below).
If the handpiece is clogged a light pressure shall be applied. The water passing through the
handpiece will dissolve completely the remaining powder inside the lumina and ensure an optimal function.