Percutaneous Leads Directions for Use
3. Remove the needle stylet from the insertion needle and verify entry into the epidural space using the
standard technique.
4. OPTIONAL. Under fluoroscopic guidance, insert the lead blank through the insertion needle and into
the epidural space. Advance the lead blank to verify entry into the epidural space, then withdraw the
blank.
5. While holding the lead stylet handle, place the steering cap over the proximal end of the stylet handle
with moderate force until it is held in place. Then slowly insert the lead, with stylet, through the
insertion needle. The lead stylet should extend to the tip of the lead.
6. OPTIONAL. If exchange of the lead stylet is desired, carefully pull out the existing stylet and insert the
preferred stylet. While inserting the stylet into the lead, if resistance is encountered, withdraw the stylet
approximately 3 cm, rotate the lead and/or stylet, and gently advance the stylet. If resistance is still
encountered, repeat the above procedure until the stylet can be fully inserted.
WARNING: Do not exchange the lead stylet while the electrode array of the lead is in the bevel of the
insertion needle. If the electrode array is in the bevel area, remove the lead from the insertion needle
before exchanging the stylet. Inserting the lead stylet in the lead while the electrode array is in the
bevel of the insertion needle increases the risk of lead and tissue damage.
WARNING: If the lead stylet is removed and reinserted, do not use excessive force when inserting the
stylet into the lead. The use of instruments, such as forceps, to grasp the stylet during insertion is not
recommended as this could result in applying excessive force and could increase the risk of lead and
tissue damage.
7. Advance the lead to the appropriate vertebral level under fluoroscopic guidance. A sufficient length
of lead (for example, at least 10 cm, or approximately three vertebrae) should reside in the epidural
space to aid in lead stabilization.
8. If using a Splitter, proceed to "Lead Connection to Splitter" in this manual. If not using a splitter,
proceed to instructions for connecting to the OR Cable assembly as follows:
SCS System
Reference Document for instructions
Precision SCS System
Precision Spinal Cord Stimulator Clinician Manual
Precision SCS System with
Precision Spinal Cord Stimulator System with
MultiWave Technology
MultiWave Technology Clinician Manual
Precision Spectra SCS
Precision Spectra System Clinician Trial Manual
System
Precision Novi SCS System
Precision Spectra Clinician Trial Manual
Percutaneous Leads Directions for Use
90970887-03 Rev A
8 of 441
Section Title
"Connecting the OR Cable Assembly"
"Connecting the OR Cable Assembly"
"Inserting the Leads into the OR Cable Connector"
"Inserting the Leads into the OR Cable Connector"