8. Grasp the tip of the tool with one hand while holding the straw in place with the other hand. Pull the
tunneling tool shaft out through the straw.
9. Push the Lead or Extension proximal ends through the straw, then withdraw the straw.
10. For Permanent Trials, proceed to "Connecting the Lead Extension" on page 9.
11. For Permanent IPG Implantation, if Extensions are used, proceed to the instructions for "Connecting
the Lead Extension" on page 9.
12. For Permanent IPG Implantation, refer to instructions for connecting to the IPG in the appropriate DFU
for your SCS system as listed in your Reference Guide.
Note: The following Codman Disposable Catheter Passers may be used in place of the Boston Scientific
tunneling tool for the 2x8 Surgical Lead only:
REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm)
Note: When using a Codman Disposable Catheter Passer, tunnel from the IPG pocket to the midline
incision using the standard technique.

Connecting the Lead Extension

1. Wipe clean the proximal end of the Lead, then insert the proximal end into the Lead Extension
connector until it stops and the retention ring (long ring) is under the Setscrew.
Note: If there appears to be an obstruction when inserting the Lead into the Lead Extension connector,
use the Hex Wrench to loosen (counterclockwise) the Setscrew and/or gently rotate the Lead to
help advance the proximal end.
2. Ensure that the Lead is fully inserted before tightening the Setscrew to prevent Lead damage.
3. Using the Hex Wrench supplied, turn the Extension connector Setscrew clockwise until it clicks,
indicating lock.
Note: • Ensure that the Hex Wrench is fully seated in the Setscrew before tightening.
• The Hex Wrench is torque-limited and cannot be overtightened.
4. Form an appropriately-sized pocket using blunt dissection on either side of midline for coiled excess
Lead and Extension connectors.
5. Place a small loop at the Lead for slack. If necessary, loosely tie a suture around the Lead-loop, but do
not tighten onto the Lead.
Surgical Leads Directions for Use
92395574-02 9 of 277