Cleaning
Before cleaning the device, remove the mains plug from the socket!
• Comply with in-house directives on hygiene
• Clean the device with a damp cloth and with non-aggressive cleaning material
• Suction jars, lids, suction bags and other accessories are to be handled in accordance
with the instruction-sheets provided
• The Medela cleaning recommendations must also be respected
Vl
Functional test
a.
Plug device in and switch on
b. Set maximum vacuum
Press knob to turn
c. Close vacuum port
d. Compare vacuum
Minimum vacuum
ok?
values not reached
Vll
Test filter
Filter open!
1. Set maximum vacuum
2. Switch on the Vario
3. Compare vacuum
ok?
Vacuum values exceed limits,
replace filter
With c/i versions,
select continuous vacuum
+ 2000 m
795 hPa
+1000 m
899 hPa
+ 500 m
955 hPa
0 m
1013 hPa
low
medium
vacuum
vacuum
M I N I M U M
V A L U E S
– 7,0 kPa
– 43 kPa
– 7,9 kPa
– 48 kPa
– 8,5 kPa
– 51 kPa
– 9,0 kPa
– 55 kPa
low
medium
vacuum
vacuum
L
I
M
I
-3 kPa
-10 kPa
Overflow protection / bacteria filter
077.0225
high
vacuum
– 59 kPa
– 66 kPa
– 70 kPa
– 75 kPa
X
high
vacuum
T
S
-10 kPa
077.0571