Therapy Log File; To Open The Therapy Log File; Patient Mode; Turn On - Medela Invia Liberty Mode D'emploi

Masquer les pouces Voir aussi pour Invia Liberty:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 91

Therapy Log file

In the Therapy Log file information regarding runtimes (on / off), pressure settings, therapy modes, alarms,
warnings and errors is listed. The last 51 events are logged and displayed.

To open the Therapy Log file

List
Graph

Patient mode

1.
To enter the Administrative mode, unlock the display, press and hold
[
], and then press [
2.
Press Selection buttons [
3.
To view additional pages, scroll with the Selection
buttons [
4.
To view the Therapy Log file as a graph, press
5.
To exit the Therapy Log file, press
The Invia Liberty pump switches automatically into main display
(in Administrative mode) 30 seconds after the last button has
been pressed. After additional 30 seconds, the pump switches
into Patient mode and the display is locked.

Turn ON

Switch Invia Liberty pump ON by pressing [
Self-test starts
If the self-test is not successful, follow instructions shown on the dis-
play or see the Alarms chapter.
When the pump is turned on in this mode, last settings (therapy
mode and pressure level) are used by default.

Check pressure

The set pressure will be shown on the display. The motor will run for a
few seconds to build up the pressure. If it runs continuously for more
than 30 seconds, check system for leaks.

Air leakage indicator

An air leakage indicator is shown on the display to visualize if there is
a)
an air leakage in the system.
a) When the indicator is "empty", the system is air tight.
].
] + [
] or [
].
"Back"
] simultaneously.
"Graph"
[
].
[
].
]
25

Publicité

Table des Matières
loading

Table des Matières