Patient Mode
XIV
1
2
3
16
EN
Switch ON Invia Liberty with [
If the self-test is not successful, see the notes for troubleshoo-
ting on the display (Chapter XVI).
The acoustic alarm function is deactivated for the first
30 sec. after turning on. During this time, the system is
evacuated and pressure is built up.
The Invia Liberty starts with the last settings used before
the pump was turned off (mode and pressure levels).
If the pump is started for the first time, the settings are
constant mode / -60 mmHg.
Check pressure
The set pressure is shown on the display. The motor will run for
a few seconds to build up the pressure. If it runs continuously
for more than 30 seconds, check system for leaks and try
again.
Standby
Switch off pressure with the "Standby" button [
- Invia Liberty changes into the standby mode
Switch off
Switch OFF Invia Liberty with [
] – self-test starts.
> 3 sec.].
> 3 sec.].