Serielle Schnittstelle; Serial Interface; Interface Sérielle - Ismatec MCP-Z Standard Mode D'emploi

Table des Matières

Publicité

Hinweis
Der Antrieb MCP-Z Standard hat beim
Einschalten eine Verzögerung von 3
Sekunden, bis die serielle Schnittstelle
zuverläßig anspricht.
Please note
When switching on the MCP-Z Standard drive,
the serial interface will only respond after a
delay of 3 seconds.
Remarque
Le moteur MCP-Z Standard a un retard de 3
secondes lors de la mise en route jusqu'à ce
que l'interface sérielle fonctionne de
manière fiable.
GND
TxD
Betriebserde
Sendedaten
Signal Ground
Transmitted Data
Prise de terre du système
Données envoyées
RxD
Empfangsdaten
Received Data
Données reçues
Kommunikationsparameter
Communication parameters
Paramètres de communication
Standardeinstellung
Default setting
Réglage standard
9600 baud, 8 bit,
1 stopbit, no parity
(DIP Switch 1 off)
Alternative
1200 baud, 8 bit,
1 stopbit, no parity
(DIP Switch 1 on)
MCP-Z STANDARD, 14-021, REV. C

Serielle Schnittstelle

RS232 IN (Eingang, weiblich)
Der Anschluss erfolgt über eine
9-polige D-Buchse.
Adressierung
Jeder Befehl beginnt mit der
Pumpenadresse (1 – 8). Sie ist
werkseitig auf 1 eingestellt und kann
mit @ verändert werden. Die Adresse
wird permanent gespeichert (d.h.
auch nach Ausschalten des Antriebs).
Die Adressierung ermöglicht die
Ansteuerung von bis zu 8 ISMATEC
Pumpen mit einer RS232-Schnittstelle.
Aufbau der Befehle
Nach der Adresse folgt ein Zeichen als
Befehl. Manche Befehle haben zusätzlich
einen Parameter, der aus 4 bzw.
5 Ziffern besteht.
Der Befehl wird mit einem carriage return
(ASCII 13) abgeschlossen. Die meisten
Befehle quittiert die Pumpe mit einem
Stern
. Ja/Nein-Abfragen beantwortet sie
*
mit + (ja) oder – (nein).
Mehrstellige Antworten beschließt die
Pumpe mit (cr) ASCII 13 und (lf) ASCII 10.
Fehlerhafte Zeichensequenzen
quittiert sie mit einem #.
Zahlenwerte gibt die Pumpe als drei-
bis fünfstellige Zahl zurück. Vier dieser
fünf Stellen sind Ziffern, eine Stelle ist
entweder ein Dezimalpunkt oder eine
führende Leerstelle.

Serial interface

RS232 IN (female)
A 9-Pin D-socket is available on the rear
panel of the pump.
Setting the pump address
Each command string begins with the
pump address (1 – 8). It is factory-set at 1
and can be changed with @.The address
will be stored permanently (even after
the pump has been switched off).The
assignment of the address enables the
user to control up to 8 ISMATEC
®
-
via one interface.
Structure of the commands
The address is followed by a character.
Some commands have an additional
parameter whichalways consists of 4
or 5 figures.
The command string is completed by
a carriage return (ASCII 13). The pump
confirms most of the commands with an
asterix
. Yes/No inquiries are answered by
*
+ (yes) or – (no).
Multi-digit replies are concluded by (cr)
ASCII 13 and (lf) ASCII 10.
Incorrect command strings are
answered by #.
Numerical values are confirmed as 3 to
5-digit figures. Four of the five digits are
numerals, one digit is either a decimal
point or a preceding blank space.
Interface sérielle
RS232 IN (entrée femelle)
Le raccordement se fait par le biais d'une
douille D à 9 pôles (femelle).
Adressage
Chaque commande commence avec
l'adresse de pompe 1 – 8. Elle est réglée
par défaut sur 1 et peut être modifiée
avec @. L'adresse est enregistrée de
manière permanente (c'est-à-dire
également après la mise hors service
drives
du moteur). L'adressage permet la
®
commande de plusieurs moteurs
ISMATEC
par le biais d'une seule et
®
même interface RS232.
Structure des commandes
Un signe suit l'adresse en tant que
commande. Certaines commandes ont
en outre un paramètre com-portant
toujours 4, respectivement 5 chiffres.
La commande se termine par un retour
à la ligne (ASCII 13). La pompe quittance
la plupart des commandes avec une
étoile
. Elle répond aux questions oui/
*
non avec + (oui) ou - (non).
La pompe termine les réponses
multipositions avec (cr) ASCII 13
et (If) ASCII 10.
Les séquences de signes erro-nées
sont quittancées avec un #.
La pompe renvoie les valeurs chiffrées
sous forme de nombre de trois à cinq
positions. Quatre de ces cinq positions
sont des chiffres, une position étant
soit un point décimal, soit un blanc en
première position.
35 of 55