The charger is capable of charging a completely depleted pack in
less than six hours in normal, room-temperature environments.
NOTE: Allow very hot or cold instruments to stabilize for one hour
at room temperature before attempting to charge.
•
Minimum and maximum ambient temperature to charge the
instrument is 10°C (50°F) and 35°C (95°F).
•
For best results, charge the instrument at room temperature.
To Charge the Instrument
•
Firmly insert the charger connector into the charge port on the
back of the instrument.
•
An LED in the battery pack is used to indicate the charge status.
• Red = charging
• Green = charged
• Amber = fault
•
If a problem is detected during charging (LED turns amber),
disconnect the charger momentarily to reset the charge cycle.
•
The charger MUST BE DISCONNECTED for the unit to operate.
2.4. Viewing Additional Pages
The Main Screen appears at instrument turn-ON.
Additional displays can be viewed by pressing the ▼ button to move to
the screen as indicated by the "soft key".
(For the monochrome display, the name of the page is displayed; for the
color display, it is represented by an icon.)
The sequence of pages are as follows and are described below:
2-9