The charger is capable of charging a completely depleted pack in
less than six hours in normal, room-temperature environments.
NOTE: Allow very hot or cold instruments to stabilize for one hour
at room temperature before attempting to charge.
Minimum and maximum ambient temperature to charge the
instrument is 10°C (50°F) and 35°C (95°F).
For best results, charge the instrument at room temperature.
To Charge the Instrument
Firmly insert the charger connector into the charge port on the
back of the instrument.
An LED in the battery pack is used to indicate the charge status.
• Red = charging
• Green = charged
• Amber = fault
If a problem is detected during charging (LED turns amber),
disconnect the charger momentarily to reset the charge cycle.
The charger MUST BE DISCONNECTED for the unit to operate.
2.4. Viewing Additional Pages
The Main Screen appears at instrument turn-ON.
Additional displays can be viewed by pressing the ▼ button to move to
the screen as indicated by the "soft key".
(For the monochrome display, the name of the page is displayed; for the
color display, it is represented by an icon.)
The sequence of pages are as follows and are described below: