Monitoring Oxygen Concentration - MSA ALTAIR 5 Manuel D'utilisation

Détecteur multigaz
Masquer les pouces Voir aussi pour ALTAIR 5:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Figure 2-6. Alarm Conditions (High Alarm shown)
If the gas concentration reaches or exceeds the alarm setpoint or the
STEL or TWA limits, the:
alarm message displays and flashes in combination with the
corresponding gas concentration:
backlight turns ON
alarm sounds (if active)
alarm LEDs flash (if active)
vibrating alarm triggers (if active).

2.7 Monitoring Oxygen Concentration

The instrument monitors oxygen concentration in ambient air. The alarm
setpoints can be set to activate on two different conditions:
Enriched - oxygen concentration > 20.8 % or
Deficient - oxygen concentration < 19.5 %.
While the instrument can detect up to 25% oxygen in the ambient air, it
is approved for use only up to 21% oxygen-content.
If an alarm activates while using the instrument, leave the
area immediately.
Remaining in the area under such circumstances can cause
serious personal injury or death.
! WARNING
2-15

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Altair 5 ir

Table des Matières