Use the Up
/ Down
arrow buttons to adjust the parameters or press the function button F1 (new) to define a new procedure. A procedure may be
duplicated with the F2 key (Duplicate). The F4 key is used to delete a selected procedure.
•
press ENTER.
•
With edit keys, enter "Device name" which is displayed in the "Reference" field in the above list.
•
Enter "Manufacturer" name which is displayed in the "Manufacturer" field in the above list.
The device is designed to operate using the "by comparison" calibration method only:
•
Select the Device Measurement mode: Channel 1 or Keypad entry.
•
Select the Reference Measurement mode: Channel 2 or Keypad entry.
•
Select the Generator: Isotech oven or Other (manual control).
Calibration procedure