THERMYS 150 G
ENERAL
A.5
Service
The unit must always be set up according to the instructions in this notice. Incomplete or poorly executed setting up may adversely affect the safety of the
operator.
The responsible authority must ensure on a regular basis that factors affecting safety do not change with time and carry out any necessary preventive work.
Before opening the unit for any work, you must ensure that all wires are disconnected from the unit
Any adjustment, maintenance or repair of an open unit must be avoided as far as possible and, if essential, must be carried out by qualified personnel who are
familiar with the risks involved.
A.5.1
Software updates
The software is updated by the UPG32 program available on the Web-site, www.aoip.com. To find out which version of firmware is installed in your unit, use the
Menu
About menu.
The quickest way to find out if an update is available is to visit the AOIP S.A.S website and look at the "Software" page.
To update the firmware, proceed as follows:
1.
If necessary, install on the PC the USB driver for communication with AOIP instruments. This driver can be downloaded from our site, along with an
information page describing the installation procedure
2.
Disconnect the leads connected to the measurement and simulation terminals
3.
Connect the instrument to the PC using the USB lead supplied with the product.
4.
Download and run the firmware update programme.
5.
Select the language then the file containing the firmware and download in the first stage.
6.
Choose the communication parameters that match the parameters of the THERMYS 150. The communication port used is a virtual port which does
not correspond to a physical port on your computer. The other parameters to be selected are defined in the diagram below.
68/107