A.8.1.1
Resistive probe simulation (temperature)
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Rt transmission function, then the appropriate "type of probe", and range using the function and navigation keys.
Select the display unit
Select measurement current (direct or pulsed)
Confirm with ENTER.
Connection is made between the two Ω terminals.
The following probes are available:
The resistive temperature probe simulation function can be used with either 2, 3 or 4 wire connection.
As for the resistance simulation function, if a polling acquisition system is used, ensure that the transmitter carries out the measurement at
least 1 ms after the current is present.
A.8.1.2
Thermocouple simulation (temperature)
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Tc transmission function, then the appropriate "type of thermocouple", using the function and navigation keys.
Select the display unit
Select the type of cold junction compensation (CSF) used. Enter the temperature of the CSF in the case of a programmed CSF.
Confirm with ENTER
Sensor
Pt 50(α = 3851)
Pt 100 (α = 3851)
Pt 100 (α = 3916)
Pt 100 (α = 3926)
Pt 200 (α = 3851)
Pt 500 (α = 3851)
Pt 1 000 (α = 3851)
Ni 100 (α = 618)
Ni 120 (α = 672)
Ni 1 000 (α = 618)
Cu 10 (α = 427)
Cu 50 (α = 428)
Simulation on Channel V2
Probe Type Caption
Pt 50
Pt 100
Pt 100-3916
Pt 100-3926
Pt 200
Pt 500
Pt 1000
Ni 100
Ni 120
Ni 1000
Cu 10
Cu 50