A.8.1.1
Resistive probe simulation (temperature)
•
Display the TRANSMISSION CONFIGURATION dialogue box:
•
Select the Rt transmission function, then the appropriate "type of probe", and range using the function and navigation keys.
•
Select the display unit
•
Select measurement current (direct or pulsed)
•
Confirm with ENTER.
Connection is made between the two Ω terminals.
The following probes are available:
The resistive temperature probe simulation function can be used with either 2, 3 or 4 wire connection.
As for the resistance simulation function, if a polling acquisition system is used, ensure that the transmitter carries out the measurement at
least 1 ms after the current is present.
A.8.1.2
Thermocouple simulation (temperature)
•
Display the TRANSMISSION CONFIGURATION dialogue box:
•
Select the Tc transmission function, then the appropriate "type of thermocouple", using the function and navigation keys.
•
Select the display unit
•
Select the type of cold junction compensation (CSF) used. Enter the temperature of the CSF in the case of a programmed CSF.
•
Confirm with ENTER
Sensor
Pt 50(α = 3851)
Pt 100 (α = 3851)
Pt 100 (α = 3916)
Pt 100 (α = 3926)
Pt 200 (α = 3851)
Pt 500 (α = 3851)
Pt 1 000 (α = 3851)
Ni 100 (α = 618)
Ni 120 (α = 672)
Ni 1 000 (α = 618)
Cu 10 (α = 427)
Cu 50 (α = 428)
Simulation on Channel V2
Probe Type Caption
Pt 50
Pt 100
Pt 100-3916
Pt 100-3926
Pt 200
Pt 500
Pt 1000
Ni 100
Ni 120
Ni 1000
Cu 10
Cu 50
A
75/107