A.7.2 Measurement By Thermocouple (Temperature); A.7.3 Thermistor Measurement (Temperature) - AOIP THERMYS 150 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Range
Resolution
Automatic wire detection
Measuring current
A.7.2
Measurement by Thermocouple (Temperature)
Display the MEASUREMENT MENU dialogue box:
Select the Tc measurement function, then the appropriate "type of thermocouple" using the function and navigation keys.
Select the display unit
Select the type of cold junction (CSF) used. Enter the temperature of the cold junction in the case of a programmed cold junction.
Confirm with ENTER.
The thermocouples available are: K, T, J, E, N, U, L, S, R, B, C, PL, Mo, NiMo/NiCo.
After a significant thermal shock, it is recommended that the unit is left for the temperature to stabilise in order to use the internal cold
junction (CSF) with maximum accuracy.
Note:
By changing the display unit to VOLT and programming CSF OFF, voltages between – 10 mV and + 75 mV can be measured.
Range
Resolution
Input impedance
A.7.3
Thermistor Measurement (Temperature)
Display MEASUREMENT MENU dialogue box:
Select Thermistor "measurement function, then select Auto, 2-Wire, 3-Wire or 4-Wire connection with the function and navigation keys.
Enter coefficients corresponding to thermistor Coeff. A, Coeff. B and Coeff. C.
Confirm with ENTER.
Wiring diagram is automatically detected by the calibrator. The THERMYS 150 displays an icon showing the connection used to carry out measurement
(
for 2 wires,
for 3 wires or
To avoid errors when measuring with 3 wires, the following is recommended:
Measure using conductors with same length, same diameter and same type.
Clean connections to avoid development of spurious electromotive force (emf).
Use Y-shaped lugs to reduce connector resistance.
Measurement on channel V1 or V2
400 ohms
(probe type: PT100)
1 mohm
2, 3 and 4 wires
0.2 to 0.45 mA
Max. 75 mV depending on the type of thermocouple used (TC
measurement function)
1 uV
> 10 MΩ
for 4 wires).
3600 ohms
(probe type: PT1000)
10 mohm
2, 3 and 4 wires
0.2 to 0.45 mA
A